Calibration system HSG-P
…allows to verify all adjustable sealing parameters (temperature, force and time) on our heat sealing devices of the type HSG.
Easy and reliable Verification of:
- Sealing force
- Sealing time
- Sealing temperature
Determining the Parameters:
The HSG-P is used to test the accuracy of the Brugger heat sealers HSG-CC, HSG-C (and HSG‑ETK). This easy-to-use tester allows you to meet the QC requirements on location in your plant. Use the HSG-P to check the values for temperature, force and time fast and efficiently–without complicated preferences. For devices of the type HSG-CC and HSG-C the displayed force can also be calibrated.
Temperature:
Simply attach the temperature jaws in place of the sealing jaws. The calibrator automatically measures six temperatures with PT-100 sensors.
Force/Time:
Simply attach the load cell to the heat protection connection. The force can be calibrated, if necessary (only HSG-CC/C). The time is displayed once the jaws reopen.
Scope of Delivery:
- Testing device HSG-P
- Power supply
- Load cell
- 1 adapter with 2 drop cables for the temperature test jaws
- 2 or 4 test jaws for the Heat-Sealing Machine
- 2 Allen screws for attaching the load cell
- 1 Socket screw key
Technical data
Compatible with | HSG-CC, HSG-C, HSG-ETK |
Dimensions device | 24 x 18 x 8 cm |
Weight set | 3 kg |
Total weight (incl. Case) | 6 kg |
Storage temperature | 0 °C – 50 °C |
Working temperature | 15 °C – 35 °C |
Relative humidity | max. 80 %, non-condensing |
Power adapter (incl.) | 110-240 V to 24V , 6 W, Europlug |
Accuracy
Measuring Type | Maximum absolute tolerance | Measuring range | Interval |
---|---|---|---|
Temperature | +/- 1.5 °C | 0 °C to 250 °C | 0,1 °C |
Force | +/- 5 N | 0 N to 1000 N | 1 N |
Time | +/- 0,03 s | 0,05 s to 99,95 s | 0,01 s |